Detail specifications can be found at Vaisala.com along with Insight software instructions. New but missing instructions in box.
HMD60 Series
The duct mounted HMD60 HUMICAP® Humidity and Temperature Transmitters are designed for monitoring humidity and temperature in demanding HVAC and light industrial applications. HMD60 series transmitters provide stable, reliable, and highly accurate (up to ±1.5%RH and ±0.1°C(0.18°F)) measurements, and are resistant to chemicals and dust.
HMD60 series transmitter options include 2 analog output models: HMD62 for measuring both humidity and temperature, and TMD62 for temperature measurements. Both HMD62 and TMD62 use 4...20 mA loop-powered current output. The HMD65 transmitter model includes both analog and digital output options.
Thanks to easy access to electronics also when the transmitter is installed to a duct, configuration and adjustment can be carried out quickly and conveniently. Available configuration and adjustment interface options range from physical trimmers and DIP switches on the transmitter\'s circuit board to Vaisala Insight PC software for Windows®.
HMD62 and TMD62 basic features and options
- Measurement options:
- HMD62 for humidity and temperature measurement
- TMD62 for temperature measurement
- Humidity parameters available as output options in HMD62: RH, Td, Tdf, A, X, Tw, H
- Analog output: 4...20mA, loop powered
- HMD62: 2 analog output channels for humidity and temperature measurements
- TMD62: 1 analog output channel for temperature measurement
- Power supply input: 10...35VDC (RL = 0 Ω)/20...35VDC (RL = 600 Ω)
- Configuration and adjustment options:
- RH and T measurement field adjustment with trimmers
- Humidity output parameter selection with DIP switches (HMD62)
- Field adjustment with MI70 handheld indicator
- Configuration and adjustment with Vaisala Insight PC software
Available parameters and default scalingHMD62 measurement parameters and default analog output scaling
Table 1shows the available output parameters and the default analog output scaling of the parameters for HMD62.
Table 1.HMD62 measurement parameters and default scalingParameterDefault scaling for 4 … 20 mA output rangeRelative humidity (RH)0…100%RHTemperature (T)-20...+80°C(-4...+176°F)Dew point temperature (Td)-40...+80°C(-40...+176°F)Dew point/frost point temperature humidity (A)0…300 g/m3(0…131.1 gr/ft3)Mixing ratio (X)0…600g/kg (0…4200 gr/lb)Wet-bulb temperature (Tw)-40...+80°C(-40...+176°F)Enthalpy (H)-40…1600kJ/kg (-9.5...695.6Btu/lb)