LogTag HAXO-8-CAL Humidity and Temperature Data Logger with Calibration Certificate
- Measurement Range: -40°C to 85°C @ 0 to 100% RH
- 2 to 3 Years of Battery Life @ 15 Minute Sampling Interval
- Customizable Alarm Settings and LED Indicators
- Designed with FDA 21 CFR Part 11 Digital Signatures
- Push Button Start with \"Pre-Start\" Fail Safe Logging
- Includes ISO 17025:2017 Humidity Calibration Certificate 25°C @ 50% RH
LogTag HAXO-8-CAL Humidity and Temperature Data Logger SpecificationsMeasurement RangeTemperature: -40°C to +85°C (-40°F to +185°F)
Humidity: 0 to 100% RH
(Limitations as detailed in Humidity Measurement Operating and Storage Conditions below.)ResolutionTemperature: <0.1°C or 0.1°F
Humidity: <0.1% RHAccuracyTemperature: < ±0.5°C @ 25°C
Humidity (0% RH to 10% RH): Better than ±6%RH
Humidity (10% RH to 80% RH): Better than ±5%RH
Humidity (80% RH to 100% RH): Better than ±7%RHData Storage Capacity8,000 Humidity & Temperature ReadingsSampling Interval Adjustable, 30 seconds to several hoursStandards ComplianceIP61Power Source3V Lithium Battery - Non-ReplaceableBattery LifeTypically 2 to 3 Years of Normal Use @ 15 Minute Sampling Interval and Monthly Data DownloadsCalibration Certificate25C @ 50% RH Humidity Calibration w/ Uncertainty Data (Included)
ISO 17025:2017 Calibration CertificateOS CompatibilityWindows 11, 10, 8 and 7 (64-Bit and 32-Bit) CompatibleStorage ConditionsTemperature: 0 to 40°C (-32°F to 140°F)
Humidity: 0 to 65% RH (Non-Condensing Environment)Weight35gCase MaterialPolycarbonateDimensions86mm x 54.5mm x 8.6mm (3.39\" x 2.14\" x 0.33\")
Humidity Measurement Operating and Storage Conditions
This chart shows the normal recommended operating range of the LogTag Humidity sensor. Conditions outside the recommended range may temporarily offset the RH signal up to ±3%RH. After return to normal conditions, the LogTag humidity sensors will slowly return towards calibration state by itself. Prolonged exposure to extreme conditions may accelerate the aging of the HAXO-8 datalogger.
Re-Conditioning Procedure
Exposure of the internal sensor to chemical vapors can interfere with the internal logging sensors and cause inaccurate readings to be recorded. In a clean environment, this will slowly be self-corrected. However, exposure to extreme conditions or chemical vapors will require the folowing reconditioning procedure to bring the internal sensor back to calibration state.
1. 80°C (176°F) at <5%RH for 36 hours (baking) followed by
2. 20-30°C (70-90°F) at >74%RH for 48 hours (re-hydration).High levels of pollutants may cause permanent damage to the internal datalogging sensors.
Exposure to Chemicals
Chemical vapors may interfere with materials used for the humidity sensor. The diffusion of chemicals into the sensor\'s polymer may cause a shift in both offset and sensitivity. In a clean environment the contaminants will slowly dissipate. The reconditioning procedure will accelerate this process. High levels of pollutants may cause permanent damage to the humidity sensor\'s polymer.