450W 110V Automatic Melt Index Instrument Lab Polymers Melt Flow Rate Tester
The automatic melt index instrument is used to measure the melt flow rate of various polymers in a viscous flow state. It is suitable for polycarbonate, polyaromatic alum, fluoroplastics and other materials with high melting temperatures. The instrument is also suitable for testing plastics with low melting temperatures, such as polyethylene, polystyrene and resin. It complies with GB/3682-2000, ASTM-D1238D3364, JIS-K7210 and ISO-1133 standards.
It is widely used in plastic raw materials products, the petrochemical industry and related universities, scientific research institutes and commodity inspection departments. It is used to determine the melt-mass flow rate (MFR).
It is applicable to the materials with the higher melting temperature, such as polycarbonate, nylon, Fluor plastics, poly oxide and other engineering plastics. It also applies to plastic materials with the lower melting temperature, such as polyethylene, polystyrene, polypropylene, ABS resin, polyformaldehyde resin.
Temperature Range Room Temperature~400˚C Temperature display error
±0.1˚C
Temperature fluctuation ±0.1˚C Recovery time ≤1min, quickly restore the constant temperature state after filling Temperature Control Resolution 0.1˚C Timing Resolution 0.01S Loading 10KG (0.325kg, 0.875kg, 0.96kg, 1.2kg, 1.64kg, 5kg) Testing methods Mass method and Volume method Scraping mode Manual and Automatic Power 450W Voltage 110V/60HZ Testing range 0.1-200g/10min Temperature Recovery Less than 3 minutes If you want more information, please feel free to contact us. Working Environment 1. Ambient temperature: 10℃-40℃
2. Ambient relative humidity: 30%-80%
3. Place the machine in the environment free of vibration and corrosive media.
4. Correctly install the machine on a solid foundation.
5. There is no strong magnetic field in the operating environment, and there is no strong convection in the air. Attention
1. The temperature of the material barrel, opening die, pressing rod, molten materials and cleaning tool is extremely high during testing and after completion. Be sure to wear high-temperature-resistant gloves, safety goggles and experimental clothing during operation.
Do not touch the high-temperature components directly with your hands.
Use specialized tools (opening die clips, pressing rod and cleaning rod) for operation.
2. The blade is very sharp. Be careful during cleaning, maintenance or adjustment.
3. Pay attention to safety when taking and placing weights, so as to prevent the weights from falling and injuring feet or damaging equipment. Handle the weights gently.
4. Ensure that the equipment is properly grounded.
5. Do not conduct internal maintenance while the device is running.
6. Do not splash the water or solvents into electrical parts during cleaning.
1. The temperature rise is quick and accurate.
2. The constant temperature precision is high.
3. After material filling, the constant temperature state can be recovered.
4. Convenient calibration and correction of experimental parameters.
5. Manual and automatic cutting test methods can be used.
6. The operation controller with the LCD display is adopted.
7. The machine is equipped with a printer, and the test results can be automatically printed out.
The operation controller is equipped with a liquid crystal display screen and control buttons to achieve clear display and convenient setting of testing parameters, the testing difficulty is greatly reduced, and operation is easy and convenient. The volumetric testing device is equipped with an induction head to achieve high-sensitivity induction, the testing result can be more accurate, and the response speed is higher. The mini printer is designed for automatically printing the data and test results, and data documents can be well kept and stored. The hidden design reduces the occupied space, and the appearance of the machine can be more attractive and compact. The large-capacity tool box is provided for holding different accessories, ensuring orderly and neat arrangement of all the tools while protecting the tools against damage and loss. Operation Method
Test Method A MFR and Method B MVR
(Note: Method B MVR can be modified (by modifying the MVR Enable parameter \"0\" or \"1\", \"0\" is off and \"1\" is on; the internal setting is already on.)
I.Test Method A MFR (When MVR is closed, use the following operation)
1. Press the \"power switch\" and press the \"TEMP-ON/OF\" heating switch.
2. Input the material testing temperature in the \"Set Temperature\" column, the number of cutting times in the \"Set Counter\" column, and the cutting interval time in the \"Set Timer\" column, and select the testing weight in the \"Countweight\" column.
3. Push the \"Opening Die Baffle\" to the left and insert the \"Opening Die\".
4. Wait for the real-time temperature displayed in \"Temperature\" to reach the set temperature and stabilize it for 15 minutes.
5. Take about 6g of the test material (put the appropriate amount according to the material), add it step by step through the funnel, and compact it gently with the \"pressing rod\". The material should be added step by step and compacted step by step. The materials are placed completely.
6. Place the \"testing rod\" and press the \"TIME\" button to start the material preheating countdown. Add the test weights after the countdown is completed.
7. The material flows out through the opening die. When the first marking of the testing rod reaches the material barrel opening, press \"RUN\" to start sampling. When the second marking of the testing rod reaches the material barrel opening, press \"STOP\" to stop the sampling test. Take valid sample strips to weigh. Take the average value and, input it into the \"Sample Weight\" column and press the \"OK\" button. The \"MFR\" column displays the test results of the mass method.
8. If you need to print data, you can move the cursor down to the \"Printer\" column and press the \"OK\" button to print the test results.
9. Pull the \"opening die baffle\" to the right, use the pressing rod to press out the \"opening die\", use the \"opening die cleaning rod\" to clean away the test material in the opening die hole, and clean the outer surface with the scraper.
10. Take a piece of 120mm long gauze and stack it into three layers. Use the \"cleaning rod\" to press the gauze into the material barrel opening and wrap it back and forth to wipe the inside of the material barrel until it is clean.
11. The above tests are completed.
12. Result calculation formula MFR: Sample average value * 600/Sampling time=g/10min
II. Test Method B MVR (When MVR is turned on, operate as follows)
1. Press the \"power switch\" and press the \"TEMP-ON/OF\" heating switch.
2. Input the material testing temperature in the \"Set Temperature\" column, the number of cutting times in the \"Set Counter\" column, and the cutting interval time in the \"Set Timer\" column, and select the testing weight in the \"Countweight\" column.
3. Push the \"opening die baffle\" to the left and insert the \"opening die\".
4. Wait for the real-time temperature displayed in \"Temperature\" to reach the set temperature and stabilize it for 15 minutes.
5. Take about 6g of the test material (according to the appropriate amount of material), add it step by step through the funnel, and compact it gently with the \"pressing rod\". The material should be added step by step and compacted step by step. The materials are placed completely.
6. Place the \"testing rod\" and press the \"TIME\" button to start the material preheating countdown. Add the test weights after the countdown is completed.
7. The material flows out through the opening die. When the first marking line of the testing rod slowly moves downwards and reaches the material barrel opening, the \"MVR Timer\" starts to calculate the fluid time. When the second marking line of the testing rod reaches the material barrel opening, the \"MVR Timer\" automatically ends calculating the fluid time and the test ends. The MVR column automatically displays the results.
8. If you need to print data, you can move the cursor down to the \"Printer\" column and press the \"OK\" button to print the test results.
9. Pull the \"opening die baffle\" to the right, use the pressing rod to press out the \"opening die\", use the \"opening die cleaning rod\" to clean away the test material in the opening die hole, and clean the outer surface with the scraper.
10. Take a piece of 120mm long gauze and stack it into three layers. Use the \"cleaning rod\" to press the gauze into the material barrel opening and wrap it back and forth to wipe the inside of the material barrel until it is clean.
11. The above tests are completed.
You can make order with no worry.
Any concern just feel free to let us know, we\'ll reply in 24 hours in normal condition.
Handling Time: Shipping is usually made in 1-3 business days after payments cleared. Orders are processed from Monday-Friday; Not on the weekends & public holidays.
Warranty: 6 months\' warranty for the machines, not for the consumables.