J-Kem Scientific 210 / Timer Temperature Controller Compact Research Grade with 100 - Hour Digital Timer, Probe / Thermocouple typeT [ blue ] ( -200 – 250° C )for Mantles upto 5 L, Accuracy 0.1° C ( 1 mL to 12 L )
Condition:Usedin Clean, Excellent working condition.
J-Kem SKU #:model-210-timer
Sigma Aldrich Sku: Z210226-1EA
Description
The same power and versatility as the Model 210, but also contains a 100-hour digital timer to turn heating OFF (or ON) at a user specified time. A new safety feature automatically disconnects power from the heater following recovery from a power failure. Contains J-KEM’s patented power control computer which provides 0.1° regulation of anything.
The temperature range desired determines the probe/thermocouple type:
- J [black] (0 – 800° C)
- K [yellow] (-50 – 1200° C)
- T [blue] (-200 – 250° C)
- RTD: (-200 – 400° C)
Specifications
Weight
5 lbs
Dimensions
6 × 6.7 × 3.25 in
Power Supply
120 VAC, 60 Hz, 10 Amps, 1200 Watts
Power Setting
80 ml toluene 100 ml flask 100 ml heating mantle SP = 80°C
50 - 500 ml (25% power)
Organic solvents heated to 50 - 110o C are set to the volume range on the front panel. When choosing between 2 power settings (i.e. 80 ml also falls within both the 10 - 100 ml range and the 50 - 500 ml range) choose the higher setting.
80 ml collidine 100 ml flask 100 ml heating mantle SP = 170°C
300 ml - 2 L (50% power)
Even though the solvent volume is less than the range of this power setting, it should be used because high temperatures require additional power.
80 ml water 100 ml flask 100 ml heating mantle SP = 80°C
10 - 100 ml (10% power)
When the heating mantle size is substantially larger (5X) than the volume being heated (i.e. the heating mantle has excess heating capacity for the volume being heated), a lower power setting gives better control.
150 ml toluene 250 ml flask 250 ml heating mantle SP = 35°C
10 - 100 ml (10% power)
Even though the solvent volume isn’t included in this power setting, it should be used because low temperatures are better regulated with less power.
Heating Liquid Volume Range:
1 - 10 mL
10 - 100 mL
50 - 500 mL
300 ml - 2 L
≥ 2 L
Approx. % of Full Power:
3
10
25
50
100

When you click on links to various merchants on this site and make a purchase, this can result in this site earning a commission. Affiliate programs and affiliations include, but are not limited to, the eBay Partner Network.