Technical Specifications
Items
Specification
Lead
Standard 12 leads
Acquisition Mode
Simultaneous 12 leads acquisition
Measurement Range
±5mVpp
Input Circuit
Floating; Protection circuit against Defibrillator effect
Input Impedance
≥50MΩ
Input Circuit Current
<0.05μA
Record Mode
Automatic: 3CH×4+1R, 3CH×4, 3CH×2+2CH×3, 3CH×2+2CH×3+1R, 6CH×2; Manual: 3CH, 2CH, 3CH+1R, 2CH+1R; Rhythm: Any lead selectable.
Filter
EMG Filter: 25 Hz / 30 Hz / 40Hz / 75 Hz / 100 Hz / 150Hz
DFT Filter: 0.05 Hz / 0.15 Hz
AC Filter: 50 Hz / 60Hz
CMRR
>100dB
Patient Leak Current
<10μA (220V-240V)
Frequency Response
0.05Hz~150Hz (-3dB)
Sensitivity
2.5mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV (error: ±5%)
Anti-baseline Drift
Automatic
Time Constant
≥3.2s
Noise Level
<15μVp-p
Paper Speed
12.5 mm/s, 25 mm/s, 50 mm/s
Recording Mode
Thermal printing system
Paper Specification
Roll 110mm x 20M
Safety Standard
IEC I/CF
Sample Rate
Normal: 1000sps/channel
Power Supply
AC: 100~240V, 50/60Hz, 30VA~100VA; DC: 14.8V/2200mAh, built-in lithium battery
Features:
- 7-inch high-resolution color touch screen
- 12-lead simultaneous acquisition and display
- ECG automatic measurement and interpretation function
- Complete digital filters, resisting baseline drift, AC, and EMG interference
- Compact and lightweight design
- Supports USB flash drive and micro SD card for extended memory
- Software upgrade via USB/SD card
- Built-in rechargeable Li-ion battery
Standard Configuration:
- Main machine: 1 PC
- Patient cable: 1 PC
- Limb electrode: 1 set (4 pcs)
- Chest electrode: 1 set (6 pcs)
- Power cable: 1 PC
- 110mm x 20M recording paper: 1 PC
- Paper axis: 1 PC