Portable ATP Fast Fluorescence Detector, Microbe Hygiene Tester, 0‑9999 RLUs, Built‑in Self‑Calibration Light Source
Note: 1.Because of the ATP Surface Test Swab needs to be frozen storage and cannot be transported in a natural state, so the ATP Surface Test Swab is NOT included in the item. Please purchase it in your local. 2.Compatible with UltraSnap ATP Surface Test Swab US2020, Don\'t support 3M ATP Surface Test Swab. Introduction
Based on the principle of firefly luminescence, the ATP fluorescence detector adopts the \"luciferase‑luciferin system\" for rapid ATP detection. All living cells contain a relatively constant amount of ATP. The ATP reading clearly reflects the quantity of microorganisms and biological residues in samples, to judge sanitary‑hygiene conditions.
This portable ATP fluorescence detector is widely used in food processing, pharmaceutical & healthcare, daily chemical industry, papermaking, industrial water treatment, environmental protection, water administration, customs quarantine and other inspection departments.
In food safety applications, it performs rapid hygiene inspection for food‑processing equipment, operator hands, processing environments and raw materials. Compact handheld design, simple operation and high accuracy, test results available within 1 minute. It is widely used by food and drug regulatory authorities.
Important Note
- ATP test swabs require frozen / refrigerated storage and cannot be shipped under normal ambient temperature. SWABS ARE NOT INCLUDED in this listing. Please purchase compatible ATP sampling swabs locally.
- This unit only supports our brand integrated liquid‑stable ATP swabs. Compatibility with third‑party swabs is NOT guaranteed.
Technical Parameters
- Imported high‑sensitivity photoelectric sensor
- Detection accuracy: 10⁻¹⁵ ~ 10⁻¹⁸ mol ATP
- Detection range: 0‑9999 RLUs
- Lower limit of detection: 1 RLU
- Power‑on self‑test: 30 seconds (built‑in self‑calibration light source)
- Detection time: 15 seconds
- Detection interference: ±5% or ±5 RLUs
- Detection mode: RLU reading
- Recorded data: RLU value, file name, record number, date, time
- Display: Large clear LCD screen
- Power supply: Built‑in 3AH high‑performance lithium‑ion battery
- Continuous working time: more than 12 hours
- Standby time: more than 300 hours
- Supports solar charger, mobile phone charger, car charger
- Reagent note: Our integrated liquid‑stable test swabs have 12‑month warranty for on‑site rapid testing
- Built‑in calibration light source for automatic qualification judgment
- USB interface: connect to PC for test‑data transmission
- Operating environment: 5℃‑40℃, relative humidity 20‑85%
- Instrument dimension (L×W×H): 190mm ×78mm ×36mm
- Instrument weight: less than 260g
Equipment Operation
- Press the button on the left side of the instrument. It runs a 30‑second calibration after power‑on, then enters the home‑page interface.
- System setting menu: allows users to configure upper‑limit, lower‑limit threshold and time parameters.
- Review historical stored test data in history menu.
ATP Bacteria Detection Steps (Surface Swab)
- Power on the unit, click detection, follow the on‑screen prompt \"put in test rod\" to enter test interface.
- Take swab out of refrigerator, leave it for 10‑15 minutes to return to room temperature.
- Remove swab from sleeve. Hold the connector, wipe target surface at 45° angle over a 10 × 10 cm area.
- Insert collector back into casing.
- Hold connector with one hand; grip capsule top with the other hand. Bend back‑and‑forth until internal valve breaks.
- Squeeze swab‑head 3 times to fully release internal liquid reagent.
- Hold swab head, flick swab downwards 5 times, immediately insert swab into instrument.
- Click detection. Test result will show after 15 seconds. Manually save test result if needed.
Test result classification:
- Test value ≤ lower limit
- Lower limit < Test value ≤ upper limit
- Test value > upper limit
Limit‑Value Setting General recommended reference thresholds (for quick hygiene monitoring, for reference only) Result Qualified Alert Unqualified RLU Reading ≤10 10‑30 >30 How to set custom threshold for your site
- Collect samples from surfaces visually clean after thorough deep cleaning.
- Repeat test 5 times and record all readings.
- Calculate average value and standard deviation of 5 groups of results.
- Set judgment thresholds based on calculation results.
Reminder: ATP reading correlates closely to microorganism quantity, but it also detects organic residues. ATP results do not have a strict one‑to‑one match with agar plate colony‑count results.
About ATP Test Swabs
Two types of swabs are available: Surface swabs and Water‑sample swabs.
- Surface swabs: for solid object‑surface hygiene test, follow above 8‑step operation.
- Water‑sample swabs: for liquid‑water sample detection.
- Take swab out of cold storage, place at room temperature for approx.10 minutes.
- Twist and remove the lower tube tip. Do not touch swab tip, avoid long‑time open‑air exposure.
- Dip swab tip into liquid sample for 1‑2 seconds and take out quickly; re‑attach sealing tube.
- Break sealing rod, squeeze to let reagent flow onto swab tip, shake swab bottom for about 2 seconds.
- Insert swab quickly into detector, press detection, read result after 15 seconds.
Swab storage specification:
- Frozen storage: valid for 5 years
- Refrigerated storage (0 ~ ‑4℃): valid for 10 months
What in Package
1 × Portable ATP Fast Fluorescence Detector1 × User Manual
Friendly Reminder ( standard terms)
- Please provide a valid delivery phone number. Without valid phone contact, safe parcel delivery cannot be guaranteed.
- Item price and shipping cost do NOT cover import duties, customs fees or local taxes. All import‑related charges shall be borne by the buyer. Please check import tariff rules of your destination country before purchasing.
- Certain countries require recipient tax ID for customs clearance (such as South Africa, Chile, Peru, Mexico, South Korea etc.). Please prepare your tax ID if your country requires it.
- If you receive defective or damaged goods, please contact us via message FIRST. We will work to resolve your issue. Kindly do not open dispute or leave negative / neutral response without prior communication. Our goal is a satisfying transaction for both buyer and seller.
When you click on links to various merchants on this site and make a purchase, this can result in this site earning a commission. Affiliate programs and affiliations include, but are not limited to, the eBay Partner Network.